our products
our products
information_top

Call us at:

(818)591-3030

USA

Click here to send us an email
tel_section

Diagnostic Automation Question & AnswerDiagnostic Automation Question & Answer Diagnostic Automation Articles Diagnostic Automation videos



Measles IgM ELISA kit

Name

Measles IgM ELISA Test

Full name

Human Measles IgM ELISA Test Kit

Category Name Infectious Disease ELISA kits
Test 96
Method ELISA method: Enzyme Linked Immunosorbent Assay
Principle Indirect ELISA: Antigen Coated Plate
Detection Range Qualitative: Positive, Weak Positive, Negative Control
Sample 10 uL
Specificity 93%
Sensitivity 93%
Total Time ~75 min
Shelf Life 12 Months from the manufacturing date

Item #:                    1409-P1   Quantity:               

Measles IgM ELISA kit

Measles_IgG_1408-1_MSDS_ELISA(05-18-2016).pdf

Measles IgM ELISA kit

callback phone number email us


Measles IgM ELISA kit description:




The Diagnostic Automation ELISA, Measles IgM is intended for use in evaluating a patient's serologic status to Measles infection.

In the Measles diagnostic test, purified Measles antigen is coated on the surface of microwells. Diluted patient serum is The Diagnostic Automation Measles IgM ELISA kit is intended for the detection and determination of Measles IgM antibodies in human sera.




Materials Provided with Measles IgG ELISA Kit:
1. Microassay plate: 96 wells coated with purified Measles virus antigen
2. Calibrator: Human serum or defibrinated plasma
3. Positive Control: Human serum or defibrinated plasma
4. Negative Control: Human serum or defibrinated plasma
5. HRP Conjugate: Ready to use, Goat anti-human IgM
6. Serum Diluent PlusI
7. Chromogen/Substrate Solution Type I
8. Wash Buffer Type I, 20X
9. Stop Solution: Ready to use,1N H2SO4 solution

Materials required but not provided:
1. Freshly distilled or deionized water
2. Dispensing system and/or pipette
3. EIA kit Microplate washer
4. EIA kit Microplate Reader with 450 single & 600-650nm dual wavelength



Measles IgM ELISA Kit Background Information:
Since the introduction of a measles virus vaccine, the U.S. has mounted an effective immunization program which has essentially eliminated measles as a major childhood disease. However, as a result of vaccine failure or the failure to be vaccinated, a recent and persistent shift in the susceptible population towards young adults has been recorded. In the case of measles, severity of illness and mortality rates are highest among adults. Thus, serology has become increasingly important as a tool for determining the immune status of the young adult population entering college or the military. In addition, the linkage between measles infection and premature delivery or spontaneous abortion supports screening pregnant mothers for susceptibility. Several diseases in addition to Measles have been associated but not causally linked to measles virus. This list includes subacute sclerosing panencephalitis (SSPE), systemic lupus erythematosus (SLE) and multiple sclerosis (MS). Patients with SSPE, a chronic degenerative neurologic disease, have documented high levels of antibody to measles virus. However, for SLE and MS there is less pronounced but statistically significant elevation in antibody levels. The significance or role of measles virus infection in these disease states is unknown at the present time. Since the presence of circulating IgG antibody to measles virus is indicative of previous infection or vaccination, screening the young adult population about to enter college or the military, pregnant women, and other individuals at risk, for seropositivity, is a valuable tool for determining their immune status. The immune response to infection or vaccination with measles virus is rapid and characteristic.
Measles specific IgM and IgG begin to appear in the circulation simultaneously. The IgM response is in 1 to 3 months, while IgG response is sustained, resulting in life-long immunity.


Measles IgM ELISA Test Principle:
Purified Measles antigen is coated on the surface of microwells. Diluted patient serum is added to wells, and the Measles IgM specific antibody, if present, binds to the antigen. All unbound materials are washed away. After adding enzyme conjugate, it binds to the antibody-antigen complex. Excess enzyme conjugate is washed off, and TMB Chromogenic Substrate is added. The enzyme conjugate catalytic reaction is stopped at a specific time. The intensity of the color generated is proportional to the amount of IgM specific antibody in the sample. The measles test kit results are read by a microwell reader compared in a parallel manner with calibrator and controls.


For more information about ELISA Kits, Rapid Tests, IFA Kits, CLIA Test Kits, or Serology tests, please see our website home page, or contact our Customer Service Representative at 818-591-3030.



Product Note:

When using DAI Measles IgM ELISA, store serum between 2 and 8 C if testing will take place within two days. If specimens  are to be kept for longer periods, store at -20 C or colder. Do not use a frost-free freezer because it may allow the specimens to go through freeze-thaw cycles and degrade antibody. Samples that are improperly stored or are subjected to multiple freeze-thaw cycles may yield erroneous results.



Got error 28 from storage engine